Cadaveric validation study demonstrates high agreement between optical navigation and CT for glenoid pin placement, suggesting its utility when CT access is limited.
Optimizing glenoid component positioning is a key determinant of stability, function and implant survival in total shoulder arthroplasty. Postoperative computed tomography (CT) is considered the reference method for assessing guide-pin or implant alignment, but it remains a retrospective evaluation and is often difficult to obtain in cadaveric research settings. Optical navigation systems such as the NDI Polaris provide immediate intraoperative geometric measurements and may represent a pragmatic alternative. This cadaveric validation study aimed to evaluate the agreement between NDI Polaris optical-navigation measurements and postoperative CT analysis for glenoid guide-pin positioning. Eight cadaveric shoulders underwent augmented-reality-guided placement of a glenoid guide-pin, followed by postoperative CT acquisition. Intraoperative optical-navigation measurements were recorded using an NDI Polaris tracking station. Two accuracy domains were analyzed relative to CT: (1) linear entry-point deviation (mm) and (2) angular deviation in version and inclination (degrees). Agreement between both measurement modalities was assessed using Bland-Altman analysis, complemented by descriptive accuracy statistics. Sixteen paired measurements were available for comparison. The mean absolute NDI-CT difference was 0.98 mm (SD 0.55; range 0.15-2.17 mm) for entry-point localization and 1.58° (SD 0.97; range 0.03-3.28°) for trajectory orientation. Bland-Altman analysis demonstrated a small systematic bias for both linear (-0.15 mm) and angular (-0.75°) parameters, with most discrepancies contained within approximately ±2 mm and ±4°, respectively. These results suggest that, under standardized acquisition conditions, optical navigation provides geometric measurements that closely approximate CT-based reference values and may serve as a practical surrogate in cadaveric research when CT access is limited. NDI Polaris optical tracking demonstrated good agreement with postoperative CT for evaluating glenoid pin positioning and may represent a practical alternative for accuracy assessment in cadaveric studies. These preliminary findings require confirmation in larger validation studies.
No takes yet. Share an insight, caveat, or question.
Letissier et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: