This is the authors' abstract. We don't add key points for this paper.
OBJECT: The authors describe the prospective use of a new hand-held point-and-shoot pupillometer (NeurOptics) to assess pupil function quantitatively. METHODS: Repetitive measurements were made in 90 pediatric participants ranging in age from 1 to 18 years, providing a total of 100 measurements under ambient light conditions. The participants consisted of 45 patients without known intracranial or ophthalmological pathological conditions as well as 45 volunteers in the outpatient setting. Quantitative pupil measurements were reliably replicated in the study participants. The mean resting pupil aperture was 4.11 mm and the minimal diameter after stimulation was 2.65 mm, resulting in a 36% change in pupil size. The mean constriction velocity was 2.34 mm/second, with a mean dilation velocity of 2.2 mm/second. CONCLUSION: Pupil symmetry was impressive in the entire cohort.