The fluorescence-based detection and counting of single protein molecules after specific binding to antibodies at interfaces is presented. A diode laser was used as the excitation source. The unspecific binding at the interface has been reduced to a level of only 0.1% of the maximum signal level. At present, the detection limit of this molecule-counting process is in the range of 10(-17) mol/L, and the dynamic range of the signal corresponds to 7 orders of magnitude of antigen concentration, but these values are not limiting. As a preliminary application in early-stage diagnosis, we have investigated the detection of a single cardiac actin molecule in human plasma, which is of interest in myocardial infarction diagnosis.
No takes yet. Share an insight, caveat, or question.
Löscher et al. (1998) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: