OBJECTIVE: To compare contrast-to-noise ratio (CNR) thresholds with visual assessment of low-contrast resolution (LCR) in filtered back projection (FBP) and iteratively reconstructed (IR) computed tomographic (CT) images. METHODS: American College of Radiology (ACR) CT accreditation phantom LCR images were acquired at CTDIvol levels of 8, 12, and 16 mGy using 2 scanner models and reconstructed using one FBP and 2 IR kernels. Acquisitions were repeated 100 times. Three board-certified medical physicists blindly reviewed the LCR section images. Pass-percentage rates (PPRs) using previous and current ACR CT accreditation criteria were compared. RESULTS: Observer PPRs for FBP images were less than 32%. For IR images, 5 of 18 settings/dose/model configurations had PPRs greater than 32% (maximum 76.3%). For CNR evaluation of FBP images, PPRs for 15 configurations were greater than 70%. For IR images, all PPRs were at least 96%. CONCLUSIONS: The CNR threshold used by the ACR CT accreditation program yields higher PPRs than visual assessment of LCR, potentially resulting in lower-quality images passing the ACR CNR criteria.