Answers are grounded in this paper and cite their sources.
A simple, accurate, and reproducible high performance liquid chromatography method for the determination of zopiclone in plasma is presented. This method involves protein precipitation with acetonitrile and reversed-phase chromatography with ultraviolet detection at 304 nm. The mobile phase was 0.067 M phosphate buffer, pH 7.95-acetonitrile (55:45, v/v). Diazepam was applied as an internal standard. The method was tested for linearity (over the range 2.5–200.0 ng/mL). The recovery (mean ± SD, n=7) was 96.76 ± 1.89%. Inter- and intra-day precision was less than 3% (n=6). The described method has been applied to the quantitative determination of zopiclone in plasma and should be useful for bioavailability investigations.
No takes yet. Share an insight, caveat, or question.
Paw et al. (2000) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: