A rapid automated method for isotopic analysis of 13C and 18O in CO2 has been developed. A variety of gas samples containing CO2 can be swept from serological tubes into a helium carrier flow; impurities are separated on a GC column so that a pure pulse of CO2 in He flows into the mass spectrometer. Isotopic ratio determinations are carried out as the pulse passes through the mass spectrometer, allowing a sample to be measured approximately every 4 min. A double, concentric needle-probe is used to flush the sample from the tube so that 100% sample recovery is achieved, maximizing sensitivity and preventing the possibility of fractionation. The precision of the technique, sigma(m-1), is better than 0.2% (0.0002 atom per cent excess) for 13C and 0.4% (0.83 p.p.m.) for 18O for 10 micromol of CO2 at natural abundance. Samples containing only atmospheric concentrations of CO2 can also be analyzed.
No takes yet. Share an insight, caveat, or question.
Prosser et al. (1991) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: