A simple, fast, reproducible (2.5% RSD at 3.0 μg/L), and sensitive method is described for quantifying As(III) (0.3 μg/L detection limit, 0.5–440 μg/L dynamic range). Anodic stripping voltammetry (ASV) is performed after accumulating arsenic at a mercury film electrode at −0.350 V vs. Ag/AgCl (saturated KCl) for 20 s in 0.2 M HCl containing 8 μM ammonium 2‐amino‐1‐cyclopentene‐1‐dithiocarboxylate (AACD), without oxygen removal. This is the first report of using AACD in ASV and in electrochemical quantification of As(III). Total arsenic is determined after sodium‐sulfite‐reduction of As(V) to As(III). Interferences are minimal. Method validation involved water and metal alloy samples.