Key result
Late right ventricular lead perforation can present with unipolar exit block and preserved bipolar capture due to anodal stimulation from the lead ring.
Case Report (n=1)
No
Late lead perforation should be considered in patients presenting with chest pain and pacemaker malfunction, where the ability to capture the myocardium during bipolar but not unipolar stimulation may indicate anodal stimulation from a perforated electrode tip.
Answers are grounded in this paper and cite their sources.
Clinicians may consider late perforation in unexplained pacemaker malfunction with differential capture; case report leaves open its diagnostic specificity.
A 75-year-old woman was referred to our hospital because of dizziness and presumed pacemaker dysfunction (Fig. 1).Eight weeks earlier, the patient had underwent implantation of a single chamber pacemaker (Sensia, Medtronic, USA) with a tined, steroid-eluting bipolar right ventricular (RV) lead (4074 CapSure Sense, Medtronic, USA) for chronic atrial fibrillation and symptomatic bradyarrhythmia.Two weeks post-implantation surgical revision had been necessary due to RV lead dislocation.Device interrogation prior to hospital discharge had been inconspicuous with a RV pacing threshold less than 1.0 V at 0.4 ms during unipolar stimulation (Fig. 2), a pacing impedance of 612 Ohms and an R wave amplitude sensing of [22 mV.Chest X-ray suggested a regular lead position in the RV apex.At presentation, the patient reported episodic dizziness and intermittent sharp left pectoral pain that was present since approximately 1 week.Intrinsic heart rate at rest was 50-60 bpm and blood pressure 130/85 mmHg.The surface Electrocardiogram (ECG) showed unipolar pacing spikes without capture (Fig. 1).Device interrogation revealed a RV pacing threshold that had increased to [6 V at 1 ms during unipolar stimulation, with a pacing impedance of 521 Ohms and an R wave amplitude sensing of [22 mV.During bipolar stimulation, a pacing threshold of 2.25 V at
No takes yet. Share an insight, caveat, or question.
Seegers et al. (2009) conducted a case report in Pacemaker dysfunction and late lead perforation (n=1). Pacemaker implantation and lead revision was evaluated. Late right ventricular lead perforation can present with unipolar exit block and preserved bipolar capture due to anodal stimulation from the lead ring.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: