A special Geiger-M\"uller counter for measuring absolute intensities of x-rays of known wave-lengths is described. The rays pass through the counter in such a way that only electrons set free in the gas of the counter are counted. The absorbing gas column is krypton at 6.31 cm pressure, 1.08 cm thick, and 0.0062 cm² in cross section. For fluorescent Zr K-rays, the absorption in the counter is 3.75 percent, as computed on the basis of the energy distribution of the radiation: Kα:Kβ:Kγ=0.829:0.154:0.017, and absorption coefficients computed from Richtmyer and Warburton's formulae. A "standard beam" of fluorescent Zr-rays from a specially constructed x-ray tube has been calibrated with the counter. With this beam, the minimum exposure of x-ray film for detectable blackening is found to be about 0.72×{}10⁶ quanta cm^-2, incident on the film, or 4.1×{}10⁴ quanta cm^-2 absorbed in the emulsions (Eastman Ultra-Speed Duplitized X-Ray Film). The total absorption coefficients of the film, celluloid, and emulsion, for the Zr-rays, are found to be, respectively, 8.63 cm^-1, 1.66 cm^-1 and 13.0 cm^-1. Time-blackening curves are given. These show discontinuities at exposures of about 10⁷ quanta cm^-2.