High-accuracy two-color heterodyne interferometry based on fundamental and second harmonic of frequency combs for air refractive index correction is developed. A monitor interferometer as well as a probe interferometer is constructed to compensate the phase noises and drifts, which are caused by introducing the acousto-optic modulators for heterodyne interferometer, to realize high-accuracy measurement of optical distance. A relative stability of 10 −10 to the total length for 500 s is achieved in the measurement of an optical path length difference between two wavelengths. In long-term measurements, the interferometric measurement results and the calculations from empirical equation of air refractive indices are in good agreement with a standard deviation of 4.1 × 10 −10 throughout the 10 h period. By applying the two-color method, high-accuracy correction of air refractive index with an uncertainty of 8.9 × 10 −8 is achieved during 10 h continuous measurements while the total refractive index changes with a range of 2.0 × 10 −6 .