We present photoelectron spectroscopy of liquid beams using a high‐resolution hemispherical photoelectron spectrometer with a multistage differential pumping system developed in our laboratory. Intense soft X‐ray synchrotron radiation (500‐1200 eV) at SPring‐8 and a relatively large detection solid angle (1.5 × 10−2 π sr) of the spectrometer enable highly efficient measurements of X‐ray photoelectron spectra of liquids. Vacuum ultraviolet photoelectron spectroscopy using the same spectrometer provides sufficiently high spectral resolution to resolve rotational isomers of gas‐phase molecules.