Key result
Continuous vectorcardiography monitoring during elective PTCA identified procedure-related myocardial infarction using maximum STC-VM with 93% sensitivity and 99% negative predictive value.
Why the study?
Does continuous on-line vectorcardiography during elective PTCA identify procedure-related myocardial infarction in patients undergoing elective PTCA?
Observational (n=192)
Does continuous on-line vectorcardiography during elective PTCA identify procedure-related myocardial infarction in patients undergoing elective PTCA?
p-value: p=<0.001
Continuous vectorcardiography monitoring during elective PTCA is a highly sensitive method for the immediate detection of procedure-related myocardial infarction.
May support real-time periprocedural MI detection during elective PTCA; leaves open need for prospective validation before practice change.
BACKGROUND: Increased creatine kinase concentrations after elective percutaneous transluminal coronary angioplasty (PTCA) have been shown to be associated with increased late cardiac mortality. OBJECTIVE: To evaluate the potential of continuous on-line vectorcardiography during elective PTCA to identify procedure-related myocardial infarction. METHODS: Patients (n = 192, ages 58 +/- 10 years), treated with elective and initially successful PTCA, were studied using vectorcardiogram (VCG) recordings. VCG monitoring was started 5 min before start of the PTCA and was carried out during the entire procedure, for at least 30 min after the first balloon inflation. ST-segment vector magnitude (ST-VM) and ST-segment change vector magnitude (STC-VM) were monitored. RESULTS: Fifteen (7.8%) procedure-related myocardial infarctions occurred. Indicators of procedure-related myocardial infarction were maximum value of ST-VM (P < 0.001) and STC-VM (P < 0.001), total ischemic time of all ST-VM episodes (P < 0.001) and STC-VM episodes (P < 0.001). The variable most closely related to a procedure-related myocardial infarction was the maximum STC-VM value during the procedure. With an optimized cutoff value, maximum STC-VM predicts a procedure-related myocardial infarction with a sensitivity of 93%, a specificity of 59% and a negative predictive value of 99%. Patients who had a stent implanted had significantly greater VCG values (P < 0.05-P < 0.001) than the group without a stent. There was a trend (P < 0.06) to a relation between increased creatine kinase concentration and stent implantation. In patients both with and without an implanted stent, greater STC-VM values were associated with procedure-related myocardial infarction (P < 0.01). CONCLUSION: Continuous VCG monitoring during elective PTCA is a promising method for immediate detection of patients at increased risk of procedure-related myocardial infarction.
No takes yet. Share an insight, caveat, or question.
Jensen et al. (2000) conducted an observational in Elective percutaneous transluminal coronary angioplasty (PTCA) (n=192). Continuous on-line vectorcardiography (VCG) was evaluated on Procedure-related myocardial infarction (p=<0.001). Continuous vectorcardiography monitoring during elective PTCA identified procedure-related myocardial infarction using maximum STC-VM with 93% sensitivity and 99% negative predictive value.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: