Answers are grounded in this paper and cite their sources.
Cryosurgery has been shown to be an effective therapy for prostate cancer. Temperature monitoring throughout the cryosurgical iceball could dramatically improve efficacy, since end temperatures of at least -40 degrees C are required. The results of this study indicate that MR thermometry based on tissue R(*)(2) has the potential to provide this information. Frozen tissue appears as a complete signal void on conventional MRI. Ultrashort echo times (TEs), achievable with half pulse excitation and a short spiral readout, allow frozen tissue to be imaged and MR characteristics to be measured. However, half pulse excitation is highly sensitive to eddy current distortions of the slice-select gradient. In this work, the effects of eddy currents on the half pulse technique are characterized and methods to overcome these effects are developed. The methods include: 1) eddy current compensated slice-select gradients, and 2) a correction for the phase shift between the first and second half excitations at the center of the slice. The effectiveness of these methods is demonstrated in R(*)(2) maps calculated within the frozen region during cryoablation.
No takes yet. Share an insight, caveat, or question.
Wansapura et al. (2001) studied this question.