We report the results of the first laser high-power measurement comparison between the Physikalisch-Technische Bundesanstalt (PTB, Germany), and the National Institute of Standards and Technology (NIST, USA). Laser power transfer standards were calibrated at both national standards laboratories between 82 W and 127 W at 1.06 µm and between 85 W and 554 W at 10.6 µm. Relative agreement between the standards of the two laboratories was demonstrated to lie between 5 × 10 −3 and 7 × 10 −3 , which is well within the combined uncertainties.