Answers are grounded in this paper and cite their sources.
The exposure rate computed from calorimeter power measurements is compared to the exposure rate of the same source (1-Ci cobalt-60) determined by means of a graphite cavity ionization chamber. A difference of 0.38% was found and is not considered significant compared to the estimated systematic uncertainty (1.3%)
No takes yet. Share an insight, caveat, or question.
Petree et al. (1975) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: