A novel carboxyl-bonded silica stationary phase was prepared by "thiol-ene" click chemistry. The resultant Thiol-Click-COOH phase was evaluated under hydrophilic interaction liquid chromatography (HILIC) mobile phase conditions. A comparison of the chromatographic performance of Thiol-Click-COOH and pure silica columns was performed according to the retention behaviors of analytes and the charged state of the stationary phases. The results indicated that the newly developed Thiol-Click-COOH column has a higher surface charge and stronger hydrophilicity than the pure silica column. Furthermore, the chromatographic behaviors of five nucleosides on the Thiol-Click-COOH phase were investigated in detail. Finally, a good separation of 13 nucleosides and bases, and four water-soluble vitamins was achieved.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2013) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: