A sensitive, selective, and quantitative method incorporating high-resolution mass spectrometry was developed for the determination of blood urea nitrogen (BUN) in bronchoalveolar lavage fluid. The method requires no sample cleanup or derivatization prior to analysis. High-performance liquid chromatography (HPLC) on a Hypersil Gold PFP column (100 × 3 mm, 3 μm particle size) connected to a C18 guard column was employed for a 10 min chromatographic separation. The detection of urea was achieved using a Thermo Scientific Q-Exactive Plus instrument incorporating selected ion monitoring (SIM) modes for the protonated adduct of urea. The urea analytical measuring range for the method is 0.047–17.134 mg/dL, resulting in a BUN analytical measurement range of 0.022–8.007 mg/dL, which allows for quantitation over 3 orders of magnitude ( R 2 = 0.999). In addition, the method is suitable for small sample volumes (15 μL) with a high level of accuracy, precision, and specificity.