A broad range of polymers have been utilized for the development of blood-contacting implantable medical devices; however, their rate of failure has raised the need for developing more hemocompatible biomaterial surfaces. In this study, a novel scaffold based on polycaprolactone incorporated with 10% and 15% (w/w) Tulbaghia violacea plant extracts were fabricated using electrospinning technique. The fabricated scaffolds were then treated with T. violacea aqueous plant extracts (100 and 1000 µg/mL) to investigate their use as interfaces for blood-contacting implants. The 10% Tvio scaffold produced the lowest mean fibre diameter (193 ± 30 nm), whereas the 15% Tvio scaffold produces the highest mean fibre diameter (538 ± 236 nm) when compared with the control polycaprolactone (275 ± 61 nm) scaffold. The number of adhered platelets was directly linked to fibre diameter and concentration of plant extract in such a way that the lowest fibre diameter scaffold (10% Tvio) inhibited platelet adhesion, whereas more platelets adhered to the scaffold with the highest fibre diameter (15% Tvio scaffolds). There was also an increase in platelet adhesion as the concentration of T. violacea was increased from 100 to 1000 µg/mL for all designed scaffolds. The improved blood compatibility demonstrated by the 10% Tvio scaffold suggests that the plant possesses antithrombogenic properties, particularly at lower concentrations.
No takes yet. Share an insight, caveat, or question.
Madike et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: