We present a new design of a portable nitrogen cryostat for operation of moving SQUIDs. A mixture of liquid and gaseous nitrogen fills a reservoir in direct contact with a copper part for the SQUID integration. The temperature at the SQUID position is 77.8 K or 78.8 K depending on orientation, and varies within /spl plusmn/10 mK during lateral movement. The cryostat can operate as a portable system for 7 hours without refilling. Washer rf SQUIDs and dc gradiometers were integrated with the cryostat. We proved the operation of the system as a moving magnetometer in an unshielded laboratory environment. Noise spectra in shielding and outside were independent of orientation. The system was equipped with a differential eddy current excitation. We show the first non-destructive material evaluation results for fatigue crack detection on stationary samples with moving SQUID sensors.
No takes yet. Share an insight, caveat, or question.
Lucı́a et al. (1997) studied this question.