Key result
Subcutaneous ICDs resulted in a similar rate of appropriate therapy compared with transvenous ICDs (19.4% vs 17.5%; P=0.45), with no difference in the total number of appropriate shocks.
Why the study?
Because S-ICDs cannot provide antitachycardia pacing for monomorphic ventricular tachycardia unlike TV-ICDs, this analysis evaluated appropriate therapy and whether antitachycardia pacing reduces appropriate shocks.
Does a subcutaneous implantable cardioverter defibrillator (S-ICD) reduce appropriate shocks compared to a transvenous implantable cardioverter defibrillator (TV-ICD) in patients with an indication for ICD therapy?
RCT (n=849)
randomized
Yes
Does a subcutaneous implantable cardioverter defibrillator (S-ICD) reduce appropriate shocks compared to a transvenous implantable cardioverter defibrillator (TV-ICD) in patients with an indication for ICD therapy?
Absolute Event Rate: 19.4% vs 17.5%
p-value: p=0.45
S-ICD provides similar shock efficacy and total number of appropriate shocks compared to TV-ICD, despite a higher proportion of patients receiving at least one shock due to the lack of antitachycardia pacing.
S-ICDs match TV-ICDs in appropriate therapy delivery; confirms comparable shock efficacy as an alternative in ICD candidates.
Background: The PRAETORIAN trial (A Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) showed noninferiority of subcutaneous implantable cardioverter defibrillator (S-ICD) compared with transvenous implantable cardioverter defibrillator (TV-ICD) with regard to inappropriate shocks and complications. In contrast to TV-ICD, S-ICD cannot provide antitachycardia pacing for monomorphic ventricular tachycardia. This prespecified secondary analysis evaluates appropriate therapy and whether antitachycardia pacing reduces the number of appropriate shocks. Methods: The PRAETORIAN trial was an international, investigator-initiated randomized trial that included patients with an indication for implantable cardioverter defibrillator (ICD) therapy. Patients with previous ventricular tachycardia <170 bpm or refractory recurrent monomorphic ventricular tachycardia were excluded. In 39 centers, 849 patients were randomized to receive an S-ICD (n=426) or TV-ICD (n=423) and were followed for a median of 49.1 months. ICD programming was mandated by protocol. Appropriate ICD therapy was defined as therapy for ventricular arrhythmias. Arrhythmias were classified as discrete episodes and storm episodes (≥3 episodes within 24 hours). Analyses were performed in the modified intention-to-treat population. Results: In the S-ICD group, 86 of 426 patients received appropriate therapy, versus 78 of 423 patients in the TV-ICD group, during a median follow-up of 52 months (48-month Kaplan-Meier estimates 19.4% and 17.5%; P =0.45). In the S-ICD group, 83 patients received at least 1 shock, versus 57 patients in the TV-ICD group (48-month Kaplan-Meier estimates 19.2% and 11.5%; P =0.02). Patients in the S-ICD group had a total of 254 shocks, compared with 228 shocks in the TV-ICD group ( P =0.68). First shock efficacy was 93.8% in the S-ICD group and 91.6% in the TV-ICD group ( P =0.40). The first antitachycardia pacing attempt successfully terminated 46% of all monomorphic ventricular tachycardias, but accelerated the arrhythmia in 9.4%. Ten patients with S-ICD experienced 13 electrical storms, versus 18 patients with TV-ICD with 19 electrical storms. Patients with appropriate therapy had an almost 2-fold increased relative risk of electrical storms in the TV-ICD group compared with the S-ICD group ( P =0.05). Conclusions: In this trial, no difference was observed in shock efficacy of S-ICD compared with TV-ICD. Although patients in the S-ICD group were more likely to receive an ICD shock, the total number of appropriate shocks was not different between the 2 groups. Registration: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT01296022.
No takes yet. Share an insight, caveat, or question.
A 2021 study conducted an RCT in indication for implantable cardioverter defibrillator (ICD) therapy (n=849). Subcutaneous implantable cardioverter defibrillator (S-ICD) vs. Transvenous implantable cardioverter defibrillator (TV-ICD) was evaluated on appropriate therapy (therapy for ventricular arrhythmias) (p=0.45). Subcutaneous ICDs resulted in a similar rate of appropriate therapy compared with transvenous ICDs (19.4% vs 17.5%; P=0.45), with no difference in the total number of appropriate shocks.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: