OBJECTIVE: To compare two commercially available kits, Japanese Encephalitis-Dengue IgM Combo ELISA (Panbio Diagnostics) and JEV-CheX IgM capture ELISA (XCyton Diagnostics Limited), to a reference standard (Universiti Malaysia Sarawak - Venture Technologies VT ELISA). METHODS: Samples were obtained from 172/192 children presenting to a site in rural India with acute encephalitis syndrome. RESULTS: Using the reference VT ELISA, infection with Japanese encephalitis virus (JEV) was confirmed in 44 (26%) patients, with central nervous system infection confirmed in 27 of these; seven patients were dengue seropositive. Of the 121 remaining patients, 37 (31%) were JEV negative and 84 (69%) were JEV unknown because timing of the last sample tested was <10 day of illness or unknown. For patient classification with XCyton, using cerebrospinal fluid alone (the recommended sample), sensitivity was 77.8% (59.2-89.4) with specificity of 97.3% (90.6-99.2). For Panbio ELISA, using serum alone (the recommended sample), sensitivity was 72.5% (57.2-83.9) with specificity of 97.5% (92.8-99.1). Using all available samples for patient classification, sensitivity and specificity were 63.6% (95% CI: 48.9-76.2) and 98.4% (94.5-99.6), respectively, for XCyton ELISA and 75.0% (59.3-85.4) and 97.7% (93.3-99.2) for Panbio ELISA. CONCLUSION: The two commercially available ELISAs had reasonable sensitivities and excellent specificities for diagnosing JEV.
No takes yet. Share an insight, caveat, or question.
Lewthwaite et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: