The retarding field energy analyzer (RFA) is a simple and reliable diagnostic technique to measure the ion temperature in the scrape-off layer and edge of magnetic fusion devices. Design and operation features of a single-sided (facing the ion flow) RFA for ion temperature measurements in the Saskatchewan Torus-Modified (STOR-M) tokamak are described. Its compact size (21 x 15 x 20 mm3) allows RFA measurements without perturbing plasma significantly. Both ion and electron temperature have been measured by RFA in the STOR-M tokamak. A method is proposed to correct the effects of ion flow on the ion temperature using the simultaneously measured Mach number. The measured electron temperature is consistent with the previously reported Langmuir probe data. Abnormal behavior of the RFA has been observed in both ion and electron modes when RFA is inserted deep into the plasma.
No takes yet. Share an insight, caveat, or question.
Dreval et al. (2009) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: