Key result
64-slice computed tomography successfully generated cardiac venous system images similar to intraoperative fluoroscopy in 74% of patients, with at least one target vein visible in 95%.
Why the study?
Can 64-slice computed tomography effectively visualize the cardiac venous system to generate images similar to intraoperative fluoroscopy for CRT planning?
Cross-Sectional (n=112)
Can 64-slice computed tomography effectively visualize the cardiac venous system to generate images similar to intraoperative fluoroscopy for CRT planning?
64-slice CT with retrospective ECG gating can effectively visualize the cardiac venous system, providing images similar to intraoperative fluoroscopy to aid in CRT planning.
Answers are grounded in this paper and cite their sources.
May aid CRT planning via venous visualization; leaves open impact on procedural success or outcomes.
BACKGROUND: Previsualization of the cardiac venous system is very important for some techniques, for example, cardiac resynchronization therapy (CRT). The aim of this study was to propose a new, user-friendly method of cardiac venous system visualization in 64-slice computed tomography (CT). METHODS: In 112 patients (66 M) aged 58 +/- 11 standard deviation, a 64-slice CT with a retrospective electrocardiogram gating was performed due to a suspicion of ischemic heart disease. Special attention was paid to the requirements for image reconstruction useful for CRT. RESULTS: In 74% of the patients, it was possible to obtain similar images to those during the CRT implantation procedure within anterior-posterior, left anterior oblique, and right anterior oblique views. The coronary sinus was clearly visible in all cases, the ostium measured 12.9 +/- 5.9 mm, and the angle of entrance 99 +/- 12 degrees. In all patients it was possible to demonstrate more than one vein; in 95%, at least one vein was clearly visible in the target area. Among the target veins, the posterolateral vein was visible most frequently (78%) in the cases as well as the lateral vein (78%). CONCLUSION: The proposed scheme in 64-slice computed tomography enables images to be generated similar to the intraoperative fluoroscopy, which can be useful in techniques where previsualization of the cardiac venous system is recommended.
No takes yet. Share an insight, caveat, or question.
Młynarski et al. (2009) conducted a cross-sectional in Suspicion of ischemic heart disease (n=112). 64-slice computed tomography with retrospective electrocardiogram gating was evaluated on Generation of images similar to those during CRT implantation procedure within anterior-posterior, left anterior oblique, and right anterior oblique views. 64-slice computed tomography successfully generated cardiac venous system images similar to intraoperative fluoroscopy in 74% of patients, with at least one target vein visible in 95%.