Key result
Switching between 1.5-mm and 1.25-mm burrs and RotaWire Floppy and Extra-support successfully allowed crossing of a severely calcified coronary lesion.
Why the study?
Forceful burr manipulation during rotational atherectomy causes severe complications, but no established strategy exists for an uncrossable calcified coronary lesion.
Case Report (n=1)
Switching between different burr sizes and wire types may be an effective strategy for crossing severely calcified coronary lesions during rotational atherectomy.
Answers are grounded in this paper and cite their sources.
May facilitate crossing resistant calcified lesions; hypothesis-generating and should not yet change practice.
Rotational atherectomy (RA) is considered to be the last resort for a severely calcified coronary artery lesion. Severe complications such as vessel perforation or burr entrapment is closely associated with forceful burr manipulation during RA. The instructions for use of Rotablator (Boston Scientific, Marlborough, MA, USA) prohibit forceful burr manipulation when rotational resistance occurs. Nevertheless, RA operators tend to forcefully manipulate the burr if it cannot cross the lesion, because there has been no established strategy for an uncrossable lesion. We present a case with a severely calcified coronary lesion, which was uncrossable by a burr 1.5 mm with RotaWire Floppy (Boston Scientific). We intentionally switched 2 burrs (1.5-mm and 1.25-mm) and 2 RotaWires (Floppy and Extra-support) to cross the lesion. Uniquely, we downsized the burr (from 1.5-mm to 1.25-mm) initially for better penetration force, and upsized the burr (from 1.25-mm to 1.5-mm) finally for better contact to the calcification within the lesion. Our case suggests that 4 different types of combinations might work in a mutually complementary manner for an uncrossable calcified lesion. < Learning Objective: In rotational atherectomy, severe complications such as vessel perforation or burr entrapment are closely associated with forceful burr manipulation. We present a case with a severely calcified coronary lesion, which was uncrossable by a burr 1.5 mm with RotaWire Floppy. We intentionally switched 2 burrs and 2 RotaWires to cross the lesion. Our case suggests that 4 different types of combinations might work in a mutually complementary manner for an uncrossable calcified lesion.>
No takes yet. Share an insight, caveat, or question.
Taniguchi et al. (2019) conducted a case report in Severely calcified coronary lesion (n=1). Intentional switch between 1.5-mm and 1.25-mm burrs and RotaWire Floppy and Extra-support was evaluated on Crossing the uncrossable calcified lesion. Switching between 1.5-mm and 1.25-mm burrs and RotaWire Floppy and Extra-support successfully allowed crossing of a severely calcified coronary lesion.