Answers are grounded in this paper and cite their sources.
A sensitive amperometric glucose biosensor based on platinum nanoparticles (PtNPs) combined aligned carbon nanotubes (ACNTs) electrode was investigated. PtNPs which can enhance the electrocatalytic activity of the electrode for electrooxidating hydrogen peroxide by enzymatic reaction were electrocrystallized on 4‐aminobenzene monolayer‐grafted ACNTs electrode by potential‐step method. These PtNPs combined ACNTs' (PtNPs/ACNTs) surfaces were characterized by scanning electron microscopy (SEM) and transmission electron microscopy (TEM). The highly dispersed PtNPs on ACNTs can be obtained. The enzyme electrode exhibits excellent response performance to glucose with linear range from 1×10−5–7×10−3 mol L−1 and fast response time within 5 s. Furthermore, this glucose biosensor also has good reproducibility. It is demonstrated that the PtNPs/ACNTs electrode with high electrocatalytic activity is a suitable basic electrode for preparing enzyme electrodes.
No takes yet. Share an insight, caveat, or question.
Zhao et al. (2007) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: