A hybrid mass spectrometer operated in low-resolution selected decomposition monitoring (SDM) mode was used for the analysis of whole fish samples for the 2,3,7,8-substituted polychlorinated dibenzo-p-dioxins and poly-chlorinated dibenzofurans (PCDD/Fs). These fish samples were previously analyzed for PCDD/Fs using a high-resolution mass spectrometer following EPA protocols. The hybrid tandem mass spectrometric method using loss of COCl gave similar quantification results to those obtained by high-resolution mass spectrometry and eliminated the interferences attributed to polychlorinated biphenyls that were encountered in the high-resolution mass spectrometric analysis. Comparison of the two methods shows that the high-resolution mass spectrometric method surpasses the tandem mass spectrometric method in other analytical figures of merit, such as detection limit, linearity and reproducibility.
No takes yet. Share an insight, caveat, or question.
Huang et al. (1991) studied this question.