Key result
Transapical implantation of a repositionable Lotus valve in the native mitral annulus resulted in acute LVOT obstruction, which was successfully managed by immediate device retrieval.
Case Report (n=1)
No
In borderline cases with narrow LVOT undergoing transcatheter mitral valve implantation, the use of a fully repositionable device is preferred to allow retrieval in case of acute LVOT obstruction.
BACKGROUND: Simultaneous transapical implantation of transcatheter heart valves in the native mitral and aortic position may be considered as an alternative to surgical valve replacement in high-risk patients presenting with combined valve disease. CASE PRESENTATION: A 59-year-old female with severe aortic stenosis, severe mitral stenosis with mild mitral insufficiency, persistent atrial fibrillation, severe chronic obstructive pulmonary disease and NYHA class of IV was evaluated by our interdisciplinary heart team. Due to the calculated Euroscore II, logistic Euroscore with 10% and 17% a decision was made towards a transapical TAVI approach. The implantation of a Sapien 3 (Edwards Lifesciences) valve in the aortic position was performed and the perioperative TEE showed a good result. The preoperative imaging revealed a narrow LVOT with risk for post interventional left ventricular outflow tract obstruction. Accordingly, it was decided against the use of balloon-expanding valves for the mitral valve position in the interdisciplinary team, as it is not repositionable. Instead, it was decided for the use of a Lotus (Boston Scientific) valve, as it is repositionable and therefore possible to retract in case of LVOT obstruction. In the present case of double valve intervention, the implantation attempt of a fully repositionable transcatheter heart valve into the native mitral annulus resulted in acute LVOT obstruction requiring immediate removal of the device. The patient was extubated and experienced uneventful postoperative recovery. CONCLUSIONS: The case shows that improved preoperative work-up is necessary for better prediction of significant LVOT obstruction following transcatheter mitral valve implantation. In borderline cases, the use of a fully repositionable device is preferred.
No takes yet. Share an insight, caveat, or question.
Heinisch et al. (2017) conducted a case report in Severe aortic and mitral stenosis (n=1). Transapical transcatheter mitral valve implantation with a repositionable Lotus valve was evaluated on Procedural success and left ventricular outflow tract (LVOT) obstruction. Transapical implantation of a repositionable Lotus valve in the native mitral annulus resulted in acute LVOT obstruction, which was successfully managed by immediate device retrieval.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: