Key result
Serial IVUS analysis of the Tryton Side Branch Stent showed no chronic stent recoil, though side branch MLA decreased from 3.5 to 2.5 mm2 at follow-up (p=0.0005) due to underexpansion and hyperplasia.
Why the study?
What are the mechanisms of restenosis and serial IVUS findings in de novo bifurcation lesions treated with the Tryton Side Branch Stent?
What are the mechanisms of restenosis and serial IVUS findings in de novo bifurcation lesions treated with the Tryton Side Branch Stent?
Serial IVUS analysis of the Tryton side branch stent reveals that reduction in lumen dimensions at follow-up is driven by side branch underexpansion and neointimal hyperplasia, not chronic stent recoil.
AIMS: To report serial intravascular ultrasound (IVUS) findings of bifurcation lesions treated with the dedicated Tryton Side Branch Stent to assess mechanisms of restenosis. METHODS AND RESULTS: The Tryton FIM study was a multicentre, prospective, single-arm, "first-in-man" study to treat de novo bifurcation lesions. Minimum lumen area (MLA) sites and overall volumes were analysed within main vessels and side branches. Overall, 27 main vessels and 22 side branches had paired baseline and follow-up IVUS. The post-intervention main vessel MLA decreased from 5.3 (4.1, 6.2) to 4.8 (3.4, 5.7) mm2 at follow-up, p=0.02, and the side branch MLA decreased from 3.5 (3.0, 3.8) to 2.5 (2.2, 3.2) mm2, p=0.0005. Stent area at the side branch did not change (mean stent area from 4.0 [3.3, 4.1] to 3.8 [3.4, 4.2] mm3/mm, p=0.95). Neointimal hyperplasia (NIH) net volume obstruction (%) measured 1.8% (0.5, 7.0) for the entire main vessel and 14.9% (2.3, 31.1) for the entire side branch stents. In both main vessel and side branches the decrease in lumen area correlated significantly with NIH. CONCLUSIONS: Serial IVUS analysis of a new side branch Tryton stent showed no chronic stent recoil. Side branch underexpansion was common and along with superimposed NIH contributed to the reduction in lumen dimensions at follow-up.
No takes yet. Share an insight, caveat, or question.
Maehara et al. (2014) studied de novo bifurcation lesions. Tryton Side Branch Stent was evaluated on Minimum lumen area (MLA) within main vessels and side branches (p=0.02). Serial IVUS analysis of the Tryton Side Branch Stent showed no chronic stent recoil, though side branch MLA decreased from 3.5 to 2.5 mm2 at follow-up (p=0.0005) due to underexpansion and hyperplasia.