Answers are grounded in this paper and cite their sources.
The objective of this study was to investigate the variation of measurements recorded when four different users of the TracerCAD system trace a model of known dimensions and volume. This complements a previous study where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the TracerCAD system by four different users. The difference in mean volume between measured results and TracerCAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, 11 out of 12 maximum volume percentage differences measured greater than 2%, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.
No takes yet. Share an insight, caveat, or question.
McGarry et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: