A novel pH sensor using a smooth Pt electrode coated with an electrically insulating polymer is addressed in this study. We present a new electrochemically modified Pt electrode. Its modification results from the anodic oxidation of pure ethylenediamine, charged with lithium trifluoromethanesulfonate, which leads to the formation of a polymer coating in one step. The assembly of the electrode surface coated with electropolymerized ethylenediamine acts as a transducer of the electrode potential vs. the pH value in aqueous solutions. It has been shown that the anodic oxidation of pure ethylenediamine yields polyethylenimine which is a high impedance polymer containing amino groups sensitive to concentration. Since the reproducibility of polymer-based pH sensors is a controversial subject, we examined both the effect of the electrode-bias time, which can be linked to the polymer thickness, and the aging of the polymer on the potentiometric response of the sensor. The pH sensor we propose here has a quasi-Nernstian behavior. It is reliable for some weeks in the pH range from 3 to 11 in aqueous media.