In order to extend the potential applications of medical ultrasonic array transducers, two optimized directions with the maximal electromechanical coefficient k33′ and minimal k31 are determined for [001] and [110] poled 0.71Pb(Mg1/3Nb2/3)O3-0.29PbTiO3 single crystals using the experimental method. The maximum values of k33′ can reach 92.8% and 93.3%, respectively, corresponding to [001]W/[110]L and [110]W/[1–11]L cuts. Furthermore, we simulate the performances of three 3.5 MHz linear array transducers based on the determined directions by PIEZOCAD. Results indicate that under the [001]W/[110]L direction, 25% broader bandwidth, 40% shorter pulse length, and 3 dB higher sensitivity can be obtained compared to the traditional Pb(Zr1−xTix)O3 transducers.