A dopamine polymer film was prepared ex situ on a bare gold template from a 10 mM dopamine solution in phosphate buffer, pH 7 followed by an overoxidation in 500 mM NaOH. The modified electrode was used for quantitative determination of epinephrine. A linear relationship between epinephrine concentration and current response was obtained in the range between 2 μM and 800 μM with the detection limit of 0.3 μM. The results have shown that using the overoxidized dopamine film it is possible to perform electrochemical analysis of epinephrine without interference of ascorbic and uric acids.
No takes yet. Share an insight, caveat, or question.
Teresa Łuczak (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: