Answers are grounded in this paper and cite their sources.
This paper describes the design, control, and preliminary assessment of a novel microprocessor-controlled multifunction ankle prosthesis that provides three microprocessor-controlled behaviors-a selectable stiffness equilibrium angle, lockable conformal damping, and swing-phase repositioning. Following a description of the motivation for providing these behaviors, the authors provide a detailed description of the device and walking controller design. This device utilizes a power-asymmetric linear actuator to provide the desired functionality in a compact and lightweight package through a combination of both hydraulic and electromechanical actuation approaches. The device is controlled for level ground walking via a finite-state machine. The functionality of the prosthesis is demonstrated by a set of benchtop experiments that characterize the ability of the prosthesis to provide the three desired behaviors and by an experiment in which the prosthesis was worn by a transtibial amputee during walking. Both sets of experiments indicate that the prosthesis provides the functionality for which it was designed.
No takes yet. Share an insight, caveat, or question.
Bartlett et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: