Answers are grounded in this paper and cite their sources.
A series of OV1-OV1701 column sets, in which the two dimensions differ in id and/or film thickness, were adopted to separate components of the volatile and semivolatile fraction of samples of plant origin. In particular two applications are presented: a target analysis approach to determine volatile suspected allergens in a medium-complexity fragrance and in sandalwood essential oil, and a fingerprint analysis approach to compare herbal extracts. Some basic chromatographic parameters of these combinations, i. e. net separation measure (S(GC x GC)), degree of orthogonality and a suitable number of modulations per peak, were also estimated by analysing two test mixtures: FAME (C4:0-C24:0) and suspected volatile allergens. Experimental results show that 0.25 mm homologous id column combination, despite their lower separation power, have good capacity to separate and resolve medium-to-complex fractions from samples of plant origin, demonstrating: (i) good system orthogonality, maximized through proper exploitation of stationary-phase selectivity and (ii) reduced 2-D column overloading effects due to the increased 2-D mass loadability, thus facilitating the analysis of mixtures whose components differ significantly in relative abundance.
No takes yet. Share an insight, caveat, or question.
Cordero et al. (2008) studied this question.