Neutron counters using He3 gas at absolute pressures up to 10 atm have been constructed and tested. The relative detection efficiency for thermal, epicadmium, and fast neutrons was measured as a function of pressure from 1 to 10 atm, and compared with similar data from a boron trifluoride counter of the same size filled to a pressure of 65 cm of B10F3 gas. The ratio He3/BF3 increases approximately linearly with pressure for epicadmium and fast neutrons, reaching values of 26.6 and 21.6, respectively, at 10 atm He3 pressure. This ratio shows a saturation effect for thermal neutrons, becoming 4.45 at 10 atm He3 pressure. The gamma-ray response has been found to be low with proper pulse shaping. High voltage plateau curves for 3- and 7-atm counters are given and indicate operating regions at least 100 V wide at a counter high voltage of about 1200 V.