A complete vacuum UV spectrophotometric device, including the source and a computerized data acquisition system, for both reflectance and transmittance measurements in the 105-200-nm wavelength range is described. The apparatus is suitable for various experimental situations such as measurements in the -200 to +100 degrees C temperature range for gases, liquids, and solids (for gases and liquids within the 0-150 bar pressure range). Drift of 0.07% per day for ambient temperature measurements and absolute reflectance as low as 0.05% are reported, illustrating the high performance reached. The setup is particularly well suited for long cycles of low temperature, high pressure experiments.
No takes yet. Share an insight, caveat, or question.
Laporte et al. (1981) studied this question.