Key result
Directional coronary atherectomy in a balloon-resistant lesion resulted in dislocation of the rotating blade and severe left main stem spasm, which was reversed by intracoronary nitroglycerine.
Case Report (n=1)
Directional coronary atherectomy passes must be performed under continuous fluoroscopic control, especially for balloon-resistant lesions, due to the unpredictability of complications like cutter dislocation.
Cutter dislocation risk in resistant lesions warrants fluoroscopic vigilance during directional atherectomy; this case report leaves open need for procedural protocol changes.
Directional coronary atherectomy (DCA) has received increased attention, especially as a bail-out procedure after failed balloon angioplasty. However, this technique may also be burdened by severe pitfalls. We report a patient with a balloon-resistant left coronary artery lesion subsequently treated with DCA. Despite its over-the-wire guidance, as the rotating cutter was advanced, it deviated from its intra-housing course and intruded into the vascular wall. Dislocation of the rotating blade was due to pressure from hard plaque tissue. After having carefully pulled back the complete catheter system, a severe spasm of the left main stem occurred, which was reversed by intracoronary nitroglycerine. The final angiography showed a left coronary artery without significant, residual stenosis. The case report underscores that DCA passes must be performed under continuous fluoroscopic control, especially for balloon-resistant lesions because of the unpredictability of DCA-imminent complication.
No takes yet. Share an insight, caveat, or question.
Bauriedel et al. (1995) conducted a case report in Balloon-resistant left coronary artery lesion (n=1). Directional coronary atherectomy (DCA) was evaluated on Procedural complications (dislocation of rotating blade, severe spasm). Directional coronary atherectomy in a balloon-resistant lesion resulted in dislocation of the rotating blade and severe left main stem spasm, which was reversed by intracoronary nitroglycerine.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: