This study was performed to assess the ablational properties of 1.4-, 1.7-, and 2.0-mm wire-guided multiple fiber catheters coupled to a XeCl excimer laser. Samples of postmortem human aorta were irradiated in blood at fluences of 40, 50, and 60 mJ/mm 2. Our results indicate that: 1) an increase in the active irradiation coverage results in an increase in the ablation efficiency; 2) the ablation efficiency is not fluence related using the 1.4- or 1.7-mm multiple fiber catheter but efficiency is fluence dependent using the 2.0-mm multiple fiber catheter; 3) the depth of tissue ablated with a multiple fiber catheter depends primarily upon the proportion of the active irradiation coverage at the catheter tip; 4) the 2.0-mm multiple fiber catheter induces craters surrounded by a larger zone of tissue damage than that observed with the 1.4- or 1.7-mm multiple fiber catheter; and 5) the 2.0-mm multiple fiber catheter should be used cautiously for laser angioplasty because of its high penetration and its risk of arterial wall damage.
No takes yet. Share an insight, caveat, or question.
Kvasnička et al. (1992) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: