An ion mobility spectrometer that has its mobility cell as a 20-segment quadrupole and functionally the q2 of a triple-quadrupole mass spectrometer has been assembled and tested. The combination of high cell pressure (maximum of 4 Torr of helium) and low axial field (20-160 V per 20.2 cm) results in negligible internal excitation of the ions despite applications of rf and axial fields. The presence of collisional focusing ensures efficient ion transmission and good sensitivity. Collision cross sections of atomic, cluster, peptide, and protein ions were measured and found comparable to literature and calculated cross sections.