Key result
A novel transputer-based physiological digital signal processing unit was capable of detecting in vitro arterial stenoses as small as 5% cross-sectional area reduction.
Why the study?
Does a novel transputer-based DSP unit accurately detect flow disturbances in in vitro models of very small arterial stenoses?
Does a novel transputer-based DSP unit accurately detect flow disturbances in in vitro models of very small arterial stenoses?
A novel transputer-based DSP unit demonstrated the capability to detect flow disturbances in in vitro models of arterial stenoses as small as 5% cross-sectional area reduction.
This paper describes the performance testing of a novel transputer-based physiological digital signal processing (DSP) unit and its application in the interpretation of pulsed Doppler ultrasound signals, obtained from models of arterial stenoses. The first test used the DSP unit as a stand-alone spectrum analyser using (1) sinusoidal frequencies (50 Hz to 10 kHz) and (2) filtered white noise (centre frequency 3 kHz, bandwidth 2.5 kHz). For the second test, the DSP unit was attached to a 30-channel multi-gate Doppler ultrasound scanner (transmitting a 4.8 MHz pulse with a repetition frequency of 4.8 kHz) and a vessel tracking unit. The Doppler ultrasound signals obtained from steady flow (100-600 ml/min) in a rigid acrylic tube (internal diameter 6 mm) were then analyzed by the DSP unit and a commercially available system. Lastly, an in vitro investigation into the flow disturbances around very small stenoses (2-25% cross-sectional area reduction), using steady flow (100-600 ml/min), was undertaken. The results indicated that the system was capable of detecting stenoses as small as 5% cross-sectional area reduction.
No takes yet. Share an insight, caveat, or question.
Robinson et al. (1996) studied arterial stenoses (in vitro models). transputer-based physiological digital signal processing (DSP) unit vs. commercially available system was evaluated on detection of flow disturbances around very small stenoses. A novel transputer-based physiological digital signal processing unit was capable of detecting in vitro arterial stenoses as small as 5% cross-sectional area reduction.