Answers are grounded in this paper and cite their sources.
A rapid-flow cytofluorometric susceptibility test for in vitro amphotericin B testing of yeasts was evaluated and compared to the National Committee for Clinical Laboratory Standards (NCCLS) M27-T reference broth macrodilution method. The flow cytofluorometric method is based on the detection of decreased green fluorescence intensity of cells stained with DiOC5(3), a membrane potential-sensitive cationic dye, after drug treatment. Testing was performed on 134 clinical isolates (Candida spp. and Torulopsis glabrata). From the dose-response curve obtained for each isolate, three endpoints were calculated by computer analysis (the concentrations at which the fluorescence intensity was reduced by 50, 80, and 90%, i.e., 50% inhibitory concentration [IC50], IC80, and IC90, respectively). A regression analysis correlating these endpoints with the M27-T MICs showed that the best agreement was obtained with IC80. The flow cytofluorometric method showed good reproducibility with control strains. These initial results suggest that the flow cytofluorometric method is a valid alternative to the NCCLS reference method.
No takes yet. Share an insight, caveat, or question.
Peyron et al. (1997) studied this question.