Key result
Temporary epicardial pacing wires post-CABG linked to cardiac arrest via undersensing-triggered R-on-T phenomenon.
Why the study?
Only a handful of cases describe temporary epicardial pacing wires triggering life-threatening events due to improper sensing in post-CABG settings.
Case Report (n=1)
Temporary epicardial pacing wires post-CABG can rarely cause life-threatening ventricular arrhythmias due to undersensing and the R-on-T phenomenon, highlighting the need to evaluate their use and turn them off if risks outweigh benefits.
Answers are grounded in this paper and cite their sources.
Undersensing by temporary epicardial wires risks R-on-T arrest post-CABG; leaves open need for revised protocols or device improvements.
Cardiac arrhythmias are common in post-coronary artery bypass graft (CABG) settings. It is a common practice to use temporary epicardial pacing wires at the end of cardiac surgery to prevent fatal arrhythmias (e.g., bradycardia,atrioventricular (AV) block, and asystole). It may also be used for sequential atrio-ventricular pacing for improved cardiac output in patients with poor ejection fraction. Epicardial wires are usually implanted around the right atrium and ventricle. Not every patient requires temporary epicardial pacing. However, certain risk factors may predispose patients to life-threatening AV blocks and ventricular tachyarrhythmias. These risk factors include advancing age, valvular surgery, poor left ventricular function, structural heart disease, diabetes mellitus, preoperative beta-blocker or digoxin use, and pre-existing history of arrhythmias. Only a handful of cases have been described in the literature where this seemingly lifesaving measure can trigger life-threatening events. Here, we describe a case where epicardial pacing wires trigger ventricular arrhythmias due to improper sensing.
No takes yet. Share an insight, caveat, or question.
Emehinola et al. (2025) conducted a case report in Post-CABG ventricular fibrillation (n=1). Temporary epicardial pacing was evaluated on Ventricular fibrillation and cardiac arrest. Temporary epicardial pacing wires caused undersensing that triggered an R-on-T phenomenon, leading to polymorphic ventricular tachycardia and cardiac arrest 48 hours post-coronary artery bypass graft.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: