Avalanche diodes which are sensitive enough to detect low energy x-rays and beta particles have been fabricated and tested. They operate at room temperature and their structure is based on large area avalanche photodiodes. X-ray response has been measured over the energy range of 4 keV to 26 keV. At 5.9 keV a FWHM of 10% was obtained. In addition, multi-element sensors have been fabricated. Beta particles were detected from a variety of isotopic sources including C1-36, Sr-90, C-14 and H-3. The tritium beta particles, with 5 keV average energy, were detected with a 20% counting efficiency from a calibrated source. Tritium beta particles from biological samples were also measured. The high sensitivity to low energy x-rays and beta particles makes this detector useful for many industrial, military and commercial applications, and in particular for biomedical research.
No takes yet. Share an insight, caveat, or question.
Squillante et al. (1986) studied this question.