Key result
Higher BMI linked to poorer 14-day ECG patch compliance driven by adhesive failures.
Why the study?
ECG patch effectiveness for AF detection is primarily limited by deteriorating skin-electrode contact rather than battery or memory constraints.
What demographic factors affect the ECG recording duration and compliance of a 14-day single-lead ECG patch in patients with atrial fibrillation?
Observational (n=256)
What demographic factors affect the ECG recording duration and compliance of a 14-day single-lead ECG patch in patients with atrial fibrillation?
Higher BMI significantly reduces the wear time and compliance of 14-day ECG patches, primarily due to adhesive failure, highlighting the need for improved adhesives in this population.
Patch ECG during health checkups detected silent AF in 5.2% aged ≥75; leaves open need for outcome trials before routine adoption.
INTRODUCTION: Atrial fibrillation (AF), affecting 3% of the US adults, is the most common arrhythmia. While ambulatory electrocardiogram (ECG) monitoring is essential for AF detection, conventional technologies have diagnostic limitations due to AF's sporadic nature. ECG patches offer extended monitoring periods, though their effectiveness is primarily limited by deteriorating skin-electrode contact rather than battery or memory constraints. OBJECTIVES: This analysis reports our experience with the Zio ECG patch (iRhythm, San Francisco, CA) in 256 AF patients. METHOD: We analyzed human and technical factors and their association with ECG recording duration using previously recorded data which employed the ECG patch as a reference. Descriptive statistics and logistic regression were used to identify associations. RESULTS: Body mass index (BMI) was found to be an independent predictor of poorer compliance in a dose-dependent manner (B = -0.046, OR = 0.955, 95% CI: 0.916-0.996, p = 0.033). Loss of adhesive was the primary reason for poor compliance (n = 25; 11%). These findings can guide researchers and clinicians in determining the appropriateness of a 14-day ECG patch based on expected wear time and patient compliance. CONCLUSION: BMI significantly impacts ECG patch compliance, primarily through adhesive failures. These findings indicate the need for improved adhesive technologies for higher BMI patients. Future device development should prioritize maintaining electrode-skin contact across diverse patient populations. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT04267133.
No takes yet. Share an insight, caveat, or question.
Dzikowicz et al. (2025) conducted an observational in Atrial fibrillation (n=256). Body mass index (BMI) was evaluated on Poorer compliance (ECG recording duration) (OR 0.955, 95% CI 0.916-0.996, p=0.033). Higher body mass index independently predicted poorer compliance with a 14-day ECG patch (OR 0.955; 95% CI 0.916-0.996; p=0.033), primarily due to adhesive failures.