Bonit is claimed to be a resorbable electrochemically deposited calcium phosphate coating consisting mainly of brushite, which is a hydroxyapatite precursor. This study involved a comparison of Ti6Al4V screw-shaped implants with and without a 15 +/- 5 microm Bonit coating in rabbit tibia and femur, after 6 and 12 weeks of insertion. The biomechanical removal torque test showed significantly increased values for the coated implants after 12 weeks (p < 0.05) but not after 6 weeks of integration. Higher bone-implant contact was found for the coated implants in the tibia after 6 weeks and for both tibial and femoral screws after 12 weeks (p < 0.05). There was no difference in the inflammatory reaction around the implants, and possible grains of the coating could be detected after 6 weeks, but not after 12 weeks of follow-up. This unloaded short-term study has shown promising results for the easily applicable and resorbable coat (Bonit) compared to uncoated titanium-alloy implants.
No takes yet. Share an insight, caveat, or question.
Reigstad et al. (2007) studied this question.