Organ doses from a set of frequent CT Examinations have been estimated from measurements in a physical anthropomorphic phantom (Remab system) by using thermoluminescence dosemeters. For the same examination techniques, organ dose coefficients (taken from the literature) obtained by Monte Carlo techniques and using mathematical phantoms. The results arrived at by the two methods are compared, trying to explain the most significant differences and their influence on the estimated values of effected dose. The experimental and calculated outcomes from such simulation are also compared to the mean dosimetric results on patients from a 1991 regional survey of CT practice in the area of Madrid. Some comments about the complementary use of information coming from both methods are made.