Answers are grounded in this paper and cite their sources.
In Denmark, haloperidol, perphenazine, and zuclopenthixol are among the most frequently requested antipsychotics for therapeutic drug monitoring. With the number of requests made at the authors' laboratory, the only rational analysis is one that can measure all three drugs simultaneously. The authors therefore decided to develop an automated high-performance liquid chromatography (HPLC) method. Two milliliters serum, 2.0 mL 10 mmol/L sodium phosphate buffer (pH 5.5), and 150 microL internal standard (trifluoperazine) solution were pipetted into HPLC vials and extracted on an ASPEC XL equipped with 1 mL (50 mg) Isolute C2 (EC) extraction columns and acetonitrile-methanol-ammonium acetate buffer (60:34:6) as extracting solution. Three hundred fifty microliters was analyzed by HPLC; a 150 x 4.6-mm S5CN Spherisorb column with a mobile phase of 10 mmol/L ammonium acetate buffer-methanol (1:9), a flow rate of 0.6-1.7 mL/min, and ultraviolet detection at 256 and 245 nm were used. Reproducibility was 5-12% and the lower limit of quantitation was 10, 1, and 5 nmol/L (4, 0.4, and 2 ng/mL) for haloperidol, perphenazine, and zuclopenthixol, respectively. The method was found to be sufficiently selective and robust for routine analysis.
No takes yet. Share an insight, caveat, or question.
Angelo et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: