Between January 1982 and July 1985, 122 cemented 'Accord' total knee replacements were performed. Out of this number, 21 were lost to follow-up because of death (in 16), revision (2) due to infection, and failure to attend (3). The post-operative alignment of 101 prostheses were compared to the incidence of radiolucencies around the tibial component at 5 years follow-up. The mechanical axis was used as a reference. The mean alignment was 0.99 degrees valgus with a standard deviation of 2.48 degrees. Some 72% of knees were within 3 degrees and 94% within 5 degrees of true alignment. Using two methods of assessing radiolucencies there was a non-significant relationship between the alignment and radiolucencies. The alignment tolerance with this prosthesis is, therefore, at least 5 degrees.