This paper reports the fabrication and testing of a helical cell separator that uses insulator-based dielectrophoresis as the driving force of its separation. The helical channel shape’s main advantage is its constant curvature radius which generates a constant electric field gradient. The presented separator was fabricated by extruding a sacrificial ink on rotating spindles using a computer-controlled robot. After being assembled, connected to the reservoir and encapsulated in epoxy resin, the ink was removed to create a helical microchannel. The resulting device was tested by circulating polystyrene microbeads of 4 and 10 μm diameter through its channel using a voltage of 900 VDC. The particles were separated with efficiencies of 94.0% and 92.5%, respectively. However, roughness in some parts of the channel and connections that had larger diameters compared to the channel created local electric field gradients which, doubtless, hindered separation. It is a promising device that could lead the way toward portable and affordable medical devices.
No takes yet. Share an insight, caveat, or question.
Guérin et al. (2014) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: