Electrochemical studies of direct orange 8 were carried out with a bare glassy carbon electrode (GCE) and a polypyrrole-coated GCE in aqueous acetonitrile medium using voltammetric techniques. One reversible couple around 0.3 V due to the redox reaction of the phenol group, one reduction peak around −0.4 V due to reduction of the azo group and one oxidation peak around 1.0 V due to oxidation of the amino group were observed. Chronocoulometric studies revealed dye adsorption on the GCE. A square-wave stripping method was developed for the determination of the dye at pH 13.0, and a linear calibration equation obtained. The reproducibility in the measurement of peak currents was confirmed from the RSD value 2.8% at 0.001 mg mL−1 concentration. A comparison of the stripping voltammetric method with the UV-Vis spectrophotometric method was made. The determination limits are wider and the RSD value is lower in the stripping voltammetric method. The concentration of the dye present in dye effluent was determined using this method.